Abstract

This study evaluated the bone incorporation process of a screw-shaped internal fixation device made of poly (L-lactide-co-D, L-lactide) (PLDLLA). Thirty-two male Wistar rats received 32 fixation devices (2 mm × 6 mm) randomly assigned to either the right or left tibia and one implant in each animal. After 7, 14, 28, and 42 days, the rats were euthanized and the specimens were subjected to microtomographic computed tomography (microCT) and histomorphometric analyses to evaluate bone interface contact (BIC%) and new bone formation (NBF%) in cortical and cancellous bone areas. The animals euthanized on days 28 and 42 were treated with calcein and alizarin red, and confocal LASER microscopy was performed to determine the mineral apposition rate (MAR). Micro-CT revealed a higher percentage of bone volume (p < 0.006), trabecular separation (p < 0.001), and BIC in the cortical (p < 0.001) and cancellous (p = 0.003) areas at 28 and 42 days than at 7 and 14 days. The cortical NBF at 42 days was greater than that at 7 and 14 days (p = 0.022). No statistically significant differences were observed in cancellous NBF or MAR at 28 and 42 days. Based on these results, it can be seen that the PLDLLA internal fixation device is biocompatible and allows new bone formation around the screw thread.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.