Abstract

This study aimed to compare the bone healing and stability of the chin with a pre-bent absorbable plate and a manually bent absorbable plate after advancement genioplasty.Patients with class II malocclusion who underwent genioplasty with bimaxillary surgery were included. After genioplasty, two absorbable bicortical screws were fixed on both sides. Then, a pre-bent absorbable plate and screws were used in the center area to fix the segment in advancement genioplasty (PB group). A manually bent absorbable plate was used for the remaining patients (MB group). Computed tomography (CT) was performed before surgery and 1 week and 1 year after surgery. Changes in the pogonion (Pog) and menton (Me) points, soft tissue points, and the ratio of bone squares under the plate were evaluated using lateral cephalometric images reconstructed with 3-dimensional CT data.32 patients were included in the study. There were no significant differences in the cephalometric measurements in the time interval from 1 week to 1 year. However, the bone square ratio in the PB group showed a significant increase after 1 year (P = 0.0021).Within the limitations of the study it seems that the use of a pre-bent absorbable plate is effective in promoting bone healing after advancement genioplasty.