Abstract

This study determined if bone conduction (BC) thresholds were influenced by vibrator type. BC thresholds were obtained for 100 subjects using mastoid placement of Radioear B-71 and B-72 and Pracitronic KH 70 bone vibrators. The nontest ear was masked (30 dB EL) and the ear ipsilateral to the vibrator was open except when testing at 3000 and 4000 Hz when the ear canal was occluded with an earplug to guard against acoustic radiation. BC thresholds at 250 Hz obtained with the B-72 and KH 70 were significantly (p less than .05) higher by 10.5 dB and at 500 Hz were significantly (p less than .05) lower by 5.5 dB than for the B-71. BC thresholds at other frequencies (1000, 2000, 3000, and 4000 Hz) were not significantly different among vibrator types. The findings indicated that reference equivalent threshold force levels (RETFLs) for BC audiometry should be specified by bone vibrator type.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call