Abstract

Resistance microwelding (RMW) is an important joining process used in the fabrication of miniature instruments, such as electrical and medical devices. The excellent corrosion resistance of 316 low-carbon vacuum melted (LVM) stainless steel (SS) wire makes it ideal for biomedical applications. The current study examines the microstructure and mechanical properties of crossed resistance microwelded 316LVM wire. Microtensile and microhardness testing was used to analyze the mechanical performance of welds, and fracture surfaces were examined using scanning electron microscopy. Finally, a bonding mechanism is proposed based on optimum joint breaking force (JBF) using metallurgical observations of weld cross sections. Moreover, comparisons with RMWs of Ni, Au-plated Ni, and SUS304 SS wire are discussed.