A previously designed capillary viscometer with measuring differential pressure was modified to measure the viscosity of non-Newtonian fluids including unadulterated blood continuously over numerous shear rates in a single measurement. Because of unavoidable experimental noise and a limited number of data, the previous capillary viscometer experienced an inaccuracy and could not directly determine a viscosity without an iterative calculation. However, in the present measurement there are numerous data available near the point of interest so that the numeric-value of the derivative,d(lnQ)/dln τω), is no longer sensitive to the method of differentiation. In addition, relatively low and wide shear rate viscosity measurements were possible because of the present precision pressure-scanning method with respect to time. For aqueous polymer solutions, excellent agreement was found between the results from the pressure-scanning capillary viscometer and those from a commercially available rotating viscometer. In addition, the pressure-scanning capillary viscometer measured the viscosity of unadulterated whole blood without adding any anticoagulants.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.