Abstract

To validate a blood-threshold (BT) segmentation software for cardiac magnetic resonance (CMR) cine images in patients with functional univentricular heart (FUH). We evaluated retrospectively 44 FUH patients aged 25±8years (mean±standard deviation). For each patient, the epicardial contour of the single ventricle was manually segmented on cine images by two readers and an automated BT algorithm was independently applied to calculate end-diastolic volume (EDV), end-systolic volume (ESV), stroke volume (SV), ejection fraction (EF), and cardiac mass (CM). Aortic flow analysis (AFA) was performed on through-plane images to obtain forward volumes and used as a benchmark. Reproducibility was tested in a subgroup of 24 randomly selected patients. Wilcoxon, Spearman, and Bland-Altman statistics were used. No significant difference was found between SV (median 57.7ml; interquartile range 47.9-75.6) and aortic forward flow (57.4ml; 48.9-80.4) (p=0.123), with a high correlation (r=0.789, p<0.001). Intra-reader reproducibility was 86% for SV segmentation, and 96% for AFA. Inter-reader reproducibility was 85 and 96%, respectively. The BT segmentation provided an accurate and reproducible assessment of heart function in FUH patients.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.