Abstract
The electrochemical behavior of oxygen (O2) in blood was studied using boron-doped diamond (BDD) electrodes. Cyclic voltammogram of O2 in a 0.1 M phosphate buffer solution solution containing 1 × 10-6 M of bovine hemoglobin exhibits a reduction peak at -1.4 V (vs Ag/AgCl). Moreover, the scan rate dependence was investigated to study the reduction reaction mechanism, which was attributable to the reduction of O2 to H2O2 via two electrons. A linear calibration curve was observed in the concentration range of 86.88-314.63 mg L-1 (R2 = 0.99) with a detection limit of 1.0 mg L-1 (S/B = 3). The analytical performance was better than those with glassy carbon or platinum electrodes as the working electrode. In addition, an application to bovine blood was performed. The O2 concentration in the blood measured on the BDD electrodes was compared to that measured using an OxyLite Pro fiber-optic oxygen sensor device. Both methods gave similar values of the O2 concentration in the range of ∼40 to 150 mmHg. This result confirms that BDD electrodes could potentially be used to detect the O2 concentration in blood.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.