Abstract

We have designed a simple, inexpensive, easy-to-build and operate apparatus for measuring blood oxygen content. The galvanic oxygen cell (fuel cell) requires as little as 1 microliter of blood and has a measuring time of 1-3 min. It is well suited for measuring oxygen content in fluids low in oxygen inasmuch as the sensitivity of the instrument is variable. Either air or water (at a known temperature and oxygen tension) can be used for calibration. No significant differences in blood oxygen content measured with our cell or the Van Slyke manometric method were found.