For the development of an implantable artificial endocrine pancreas, a sensor for blood glucose measurement is needed providing a long-term stability. This goal can be achieved by the application of infrared spectroscopy which, unlike electrochemical sensors, responds directly to the glucose molecule. An investigation under physiological conditions revealed five glucose absorption bands in the near and middle infrared range. These are 1040, 1085, 1109, 1160 and 1365 cm-1. Only the 1040 cm-1 frequency coincides with none of the other infrared-active blood substances like proteins, lipids and urea. Nevertheless, the other absorption bands too, especially the 1109 cm-1 frequency, can be used for blood glucose measurement, if the superimposed absorptions are compensated. Methods for the compensation have been found. Technically feasible embodiments of an infrared glucose sensor are described.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.