Abstract
Spiral-coiled membrane lungs of 0.6 m2 surface area were fabricated from silicone rubber membranes coated with four types of commercially available pure-gum polysiloxane rubber: 1) polytrifluoropropylmethylvinylsiloxane, 2) polymethylphenylsiloxane, 3) polymethylvinylsiloxane, and 4) polydimethylsiloxane. Membranes lungs were tested in an arteriovenous shunt without added heparin after bypass was begun. All group 4 had major rise in resistance to blood flow or clotting in over 33% of the membrane lungs studied, after 6 hr of bypass. Membrane lungs with polydimethylsiloxane (group 4) had no rise in resistance to blood flow during 24 hr of bypass and had the least amount of changes in blood platelet count.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.