Abstract

The limit of detection of gold by electrothermal atomisation and double-resonance laser-induced atomic fluorescence can reach the low femtogram level as absolute amount of gold deposited in the furnace. At these high sensitivities, even dry firings of the graphite tube may result in a signal, which is unequivocally attributed to atomic fluorescence of the gold content of the tube released during atomisation. The occurrence and the analytical relevance of such blank signals are presented and discussed in this note.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.