Abstract

Background: The aim of this study was to evaluate the variation in the daily output of a linac Varian Edge TrueBeam over a month, using a Daily QA3 (QA3) and machine performance check (MPC), after Task Group 51protocol and baseline data had been established. Methods and Materials: The daily output data were collected for all photons (6, 10 MV, and 6, 10 FFF) and electrons (6, 9, 12, and 15 MeV) energies. A Bland-Altman-Plot and intraclass correlation coefficients were used to assess agreement and reliability. Results: MPC data showed small random deviations of ≤ 2% and agreed with the QA3 measurements. The maximum mean absolute difference between the QA3 and MPC data was observed with 6 MeV (0.5633), with limits of agreement of (−0.5392–1.6658). Conclusion: These results suggest that MPC data closely track QA3 data and that the two forms are interchangeable

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.