Abstract

BackgroundTo quantify image quality and radiation doses in regions adjacent to and distant from bismuth shields in computed tomography (CT). MethodsAn American College of Radiology accreditation phantom with four solid rods embedded in a water-like background was scanned to verify CT number (CTN) accuracy when using bismuth shields. CTNs, image noise, and contrast-to-noise ratios (CNRs) were determined in the phantom at 80–140 kVp. Image quality was investigated on image portions in the zones adjacent (A zone) to and distant (D zone) from a bismuth shield. Surface radiation doses were measured using thermoluminescent dosimeters. Streak artefacts were graded on a 3-point-scale. ResultsChanges in CTN caused by a bismuth shield resulted in changes in X-ray spectra. CTN changes were more apparent in the A zone than in the D zone, particularly for a low tube voltage. The degrees of CTN changes and image noise were proportional to the thickness of the bismuth shields. A 1-ply bismuth shield reduced surface radiation doses by 7.2%–15.5%. The overall CNRs were slightly degraded, and streak artefacts were acceptable. ConclusionsUsing a bismuth shield could result in significant CTN changes and perceivable artefacts, particularly for a superficial organ close to the shield, and is not recommended for quantification CT examinations or follow-up CT examinations.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call