Abstract

Lead aprons are widely used in medical imaging to protect radiographers and patients from harmful radiation. However, lead is not a desirable material for use in wearable radiation protection due to its heavy weight, inflexibility, poor durability, and toxicity. In this study, we explored the suitability of bismuth oxide (Bi2O3) coating for textiles as an alternative to lead. The intention was to demonstrate the concept and technology that will achieve a lead-equivalent lightweight X-ray protective textile material with improved wearability. The primary objective was to evaluate the X-ray shielding efficiency of two textile materials coated with Bi2O3. To do so, X-ray exposures were made at the system setting of 80 kVp, 12 mAs, and 80 SID (the distance from the X-ray beam source to the specimen). It is evident from this study that Bi2O3 in a suitable resin matrix can be coated on fabrics and is an effective method to produce flexible, wearable, and lead-free aprons. Coated polyester fabrics with over 50% Bi2O3 showed enhanced shielding ability for transmitted X-rays. This research has shown that microparticle size Bi2O3 can be effective for X-ray attenuation.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.