Abstract

A straightforward protocol for measuring low concentration levels of Ag(I) using bismuth film electrode (BiFE) in combination with anodic stripping voltammetry was developed and studied as a suitable alternative to current methods for detecting the dissolution of silver nanoparticles (Ag NPs). The BiFE was prepared in-situ on the surface of a supporting glassy carbon electrode. Several key operational parameters were examined and optimized, e.g. accumulation potential, accumulation time and concentration of Bi(III). The BiFE revealed good linear operation in the examined concentration range of 10–90μgL−1 Ag(I) with r2 of 0.998, calculated limit of detection (3σ) of 2.1μgL−1 obtained after 60s accumulation, and favorable repeatability with RSD of 3.6% (50μgL−1, n=10). The method was successfully tested through measuring Ag(I) in the supernatant of a commercial Ag NP suspension and the performance was superior to the potentiometric measurement using silver ion-selective electrode.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.