Abstract

A new reagentless biosensor for differential pulse voltammetric detection of hydroquinone in the wastewater of the photographic process was developed. Crude extracts of several vegetables, such as arracacha, cassava, corn, ginger, guava, pineapple, swiss chard, and sweetsop were obtained as sources of peroxidase (PER; EC 1.11.1.7). Among them, a ginger crude extract was selected due to the highest specific activity observed after its immobilization by cross‐linking with glutaraldehyde in a prepared chitosan matrix that was incorporated in a carbon paste electrode. In the presence of hydrogen peroxide, this enzyme catalyzes the oxidation of hydroquinone to o‐quinone. The recovery of hydroquinone from the samples ranged from 98.4–105%, and a rectilinear analytical curve for hydroquinone concentration from 2.50 × 10−4 to 2.40 × 10−3 mol L−1 (r = 0.997) was obtained. The detection limit was 2.50 × 10−5 mol L −1, and the relative standard deviation was lower than 1.2% for 4.95 × 10−4 mol L−1 hydroquinone in 0.1 mol L−1 phosphate buffer solution at pH 7.0 (n = 10).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.