Abstract
Objective To verify the feasibility and the accuracy for the biopsy of targets not visible on sonography but shown on CT in a phantom by using real-time virtual navigation system(RVS).Methods The tissue mimicking phantom,made of gel,contained 30 red spheres which were 10 mm in diameter,randomly placed.These spheres were invisible on sonography,but visible on CT.A junior operator and a senior operator performed the biopsy respectively,and each sphere was biopsied once guide by both RVS and conventional ultrasound for each operator.The red dye in the biopsied sample meant successful performance.The success rate of biopsy and the length of the red dyed sample were recorded.Results There was no statistically difference between the performance results of the two operators.The success rate and the median length of red dyed sample biopsy of RVS were respectively 96.7% (58/60) and 7.5 mm (rang,6-9 mm) which were statistically higher and longer than that of conventional ultrasound,43.3%(26/60) and 0 mm(rang,0 - 5 mm) ( P <0.05).Conclusions Biopsy guided by RVS for the target invisible to ultrasound is a feasible and accuracy method,which will be a new guided method for interventional ultrasound. Key words: Ultrasonography ; Phantoms, imaging; Biopsy, needle; Real-time virtual navigation system
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.