A multichannel biopotential amplifier for simultaneous use with biomagnetic measurements in a magnetically shielded room is designed and evaluated. Particular care is taken to make the amplifier electromagnetically compatible with the biomagnetic instruments over the whole frequency spectrum, from DC to RF. The electromagnetically quiet environment allows the use of high electrode impedances; the preamplifier has been designed accordingly. Special care is taken to analyse the coupling mechanisms of mains interference to the amplifier. Over 170 simultaneous electric and magnetic recordings have been performed using the system in a hospital environment.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.