Abstract
Catheter ablation by radiofrequency (RF) energy has successfully eliminated cardiac tachyarrhythmias. RF ablation lesions are created by thermal energy. Electrode catheters with 4-mm-tips have been adequate to ablate arrhythmias located near the endocardium; however, the 4-mm-tip electrode does not readily ablate deeper tachyarrhythmia substrate. With 8- and 10-mm-tip RF electrodes, ablation lesions were larger; yet, these catheters are associated with increased risk for coagulum, char and thrombus formation, as well as myocardial steam rupture. Cooled-tip catheter technology was designed to cool the electrode tip, prevent excessive temperatures at the electrode tip–tissue interface, and thus allow continued delivery of RF current into the surrounding tissue. This ablation system creates larger and deeper ablation lesions and minimizes steam pops and thrombus formation. The purpose of this article is to review cooled-tip RF ablation biophysics and outcomes of clinical studies as well as to discuss future technological improvements.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.