Abstract

BackgroundBiometric parameters play an important role in studies on myopia. This study aimed to explore the application of a commercially available two-dimensional visualized swept-source optical coherence tomography (OCT) system, for in vivo biometric measurement in multiple myopia model species. MethodsIn this study, chickens, guinea pigs, and C57BL/6 mice underwent eye imaging with the commercially available OCT (CASIA2), and the original images were used to calculate the central corneal thickness, anterior chamber depth (ACD), lens thickness (LT), vitreous chamber depth (VCD), and axial length (AL). The retinal thickness and choroidal thickness were also calculated in chicken eyes. The repeatability of the biometric measurement outcomes was analyzed. ResultsExcellent repeatable AL measurements were obtained for all three species, with an intraclass correlation coefficient (ICC) of ≥0.941 and a within-subject standard deviation of ≤0.055. Excellent repeatability was found in chicken eyes for ACD, LT, and VCD, with an ICC of ≥0.932; in guinea pig eyes for ACD and VCD, with an ICC of ≥0.934; and in mouse eyes for LT, with an ICC of ≥0.941. ConclusionsIt is effective to use commercially available OCT to measure biometric parameters in chickens, guinea pigs, and C57BL/6 mice. This methodology could potentially increase the accuracy and efficiency of future myopia animal experiments.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.