Abstract

We tested the hypothesis that a pedicle screw construct's height is an important factor in strengthening a screw-rod system. Six corpectomy constructs were made, each using two ultra-high-molecular-weight polyethylene blocks, 6.5-mm pedicle screws, and two 6.35-mm rods. Pedicle screws were placed at +10-, +5-, 0-, and -5-mm depths in relation to the dorsal surface of the corpectomy model. Nondestructive testing was performed in flexion/extension and in torsion. For all modes tested, the screw-rod constructs continued to increase in stiffness as the height of the construct was lowered, and this was statistically significant at all heights tested (P < 0.001). The stiffness increased 232% when comparing flexion at +10 and -5 mm and increased 231% in extension from +10 to -5 mm. The torsional stiffness increased 171% when comparing +10 and -5 mm. Thus, lower-profile instrumentation systems should be used to take advantage of this by decreasing the size and bulkiness of the implants while increasing the strength of the construct.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call