Abstract

The objective was to compare L4/5 range of motions of fusion constructs using anchored cages. Twelve human cadaveric spine were tested in intact condition, and divided into TLIF and PLIF groups. Testing consisted in applying pure moments in flexion-extension, lateral bending and axial rotation. The computation of intersegmental motion was assessed using 3 D biplanar radiographs. In TLIF group, the addition of contralateral transfacet decreased flexion-extension motion (39%; p = 0.036) but without difference with the ipsilateral pedicle screw construction (53%; p = 0.2). In PLIF group, the addition of interspinous anchor reduced flexion-extension motion (12%; p = 0.036) but without difference with the bilateral pedicle screw construction (17%; p = 0.8).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.