Abstract
The aim of this study was to determine the influence of two different artificial nucleus implants for lumbar discs on the biomechanical properties of the spinal motion segment. From 7 calf spines, the lumbar segments L2/3 and L4/5 were harvested, the lamina and the muscles were removed with care to preserve the anterior and posterior longitudinal ligament. The segments were fixed on a special testing device mounted in an universal testing machine (Zwick, Ulm, Germany). Physiological load with a maximum of 7 Nm was applied in flexion/extension, right and left side bending, and right torsion. Three cycles of measurement were performed in all specimen: a) the intact segment, b) segment after nucleotomy, c) 6 segment with 2 PDN devices (Ray Medica) and 8 segments with an autopolymerized polymer. There was a statistically significant (p = 0.0078/p = 0.0313) increase in segmental mobility in all directions after nucleotomy. After introduction of two PDN implants as well as the polymer implant, there was a restoration of segmental mobility for all movement directions with no statistically significant difference (p > 0.05) compared to the intact situation before nucleotomy.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.