Abstract

Internal distraction devices are commonly used in congenital micrognathia. The eventual need for device and screw removal can be challenging, requiring extensive dissection and disturbance of bone regenerate. Bioabsorbable poly-L-lactide (PLLA) screws, compared to traditional titanium screws, simplify device removal. Previous in vivo studies have found that the maximal compressive force generated by mandibular distraction is 69.4N. We hypothesized that PLLA screws could support these compressive/distraction forces. Ten mandibles were obtained from 5 canine cadavers. Paired mandibles from the same cadaver were each fixated to a mandibular distractor with eight screws (either titanium or PLLA). Devices were each set to 15 and 30 mm of distraction distance. Compression force of 80 N was then generated parallel to the axis of the distraction device. Distractor displacement was measured to detect any mechanical failure during this pre-set load. Finally, if no failure was observed at 80 N, a load-to-failure compression test was done in the PLLA group to determine the mechanical failure point. All distractors in both the titanium and PLLA screw groups withstood 80 N of compression without failure. When the load-to-failure test was performed in the PLLA group, the average device failure point was 172.8 N (range 148-196 N). Review of high-frame-rate video demonstrated that all failures occurred due to the PLLA screws breaking or falling out. Bioabsorbable PLLA screws can withstand compressive forces more than double that of the maximal in vivo forces needed during mandibular distraction. These screws may be an acceptable alternative for the fixation of internal mandibular distractors.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call