Abstract

Subtalar dislocation is a rare injury caused by high-energy trauma. Current treatment strategies include leg casts, internal fixation and external fixation. Among these, external fixators are the most commonly used as this method is believed to provide better stabilization. However, the biomechanical stability provided by these fixators has not been demonstrated. This biomechanical study compares two commonly used external fixators, i.e. Mitkovic and Delta. CT imaging data were used to reconstruct three-dimensional models of the tibia, fibula, talus, calcaneus, navicular, cuboid, three cuneiforms and five metatarsal bones. The 3D models of the bones and cartilages were then converted into four-noded linear tetrahedral elements, whilst the ligaments were modelled with linear spring elements. Bones and cartilage were idealized as homogeneous, isotropic and linear. To simulate loading during walking, axial loading (70N during the swing and 350N during the stance phase) was applied at the end of diaphyseal tibia. The results demonstrate that the Mitkovic fixator produced greater displacement (peak 3.0mm and 15.6mm) compared to the Delta fixator (peak 0.8mm and 3.9mm), in both the swing and stance phase, respectively. This study demonstrates that the Delta external fixator provides superior stability over the Mitkovic fixator. The Delta fixator may be more effective in treating subtalar dislocation.