Abstract

Understanding of implant failure mechanisms is important in the successful utilization of anterior cervical plates. Many variables influence screw purchase, including the quality of the bone. The purpose of this study was to assess the relationship of screw pullout and screw insertional torque across a wide range of bone mineral densities (BMDs). A total of 54 cervical vertebrae in 12 cervical spines were evaluated for BMD using dual-energy x-ray absorptiometry scanning. Actual and perceived peak torques of 3.5-mm anterior cervical screws were determined at each level followed by screw pullout strength testing. A high correlation was observed between screw pullout strength and BMD. However, there was a low correlation of peak insertional torque to pullout strength. These findings suggest the quality of the bone is more instrumental in the success or failure of anterior cervical screws than is the insertional torque with which the screws are placed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.