Abstract

The percutaneous knotless repair technique for Achilles tendon ruptures utilizes a Percutaneous Achilles Repair System (PARS) device for suturing the proximal tendon and 2 suture anchors for fixing the sutures into the calcaneus. Determining the best position of the suture anchors may optimize the strength of this repair. Twelve pairs of human ankle cadaveric specimens were randomly assigned to receive suture anchors placed at 45°, 90°, or 135° from the sagittal plane. The anchors were tensioned according to a protocol representing progressive, postoperative rehabilitation. Load, number of loading cycles, displacement, and mode of failure were recorded. With the anchors placed at 45°, 90°, and 135°, the ultimate failure loads were mean 265 ± 64 N, 264 ± 75 N, and 279 ± 40 N, and the total number of loading cycles were mean 459 ± 166, 466 ± 158, and 469 ± 110, respectively. The effect of anchor angle on failure load, number of loading cycles, and displacement was not statistically significant. Visually, the anchors at 45° and 90° demonstrated sutures cutting through the bone. We found no statistically significant difference in pullout strength between the 3 different anchor angles. Sutures cutting through the bone may be a concern with acute anchor angles. This suggests that a 135° anchor angle may result in a lower risk of tendon elongation with the percutaneous knotless repair technique. Cadaveric laboratory study.