Abstract

To compare the biomechanical characteristics of self-made nickel-titanium shape memory alloy stepped plate with calcaneal plate and cannulated compression screws in fixing calcaneal osteotomy. Calcaneal osteotomy was operated on 6 fresh-frozen lower limbs collected from donors. Then three kinds of fixation materials were applied in random, including the self-made nickel-titanium shape memory alloy stepped plate (group A), calcaneal plate (group B), and cannulated compression screws (group C). Immediately after fixation, axial loading of 20-600 N and 20 N/s in speed was introduced to record the biomechanical data including maximum displacement, elastic displacement, and maximum load. Then fatigue test was performed (5 Hz in frequency and repeat 3 000 times) and the same axial loading was introduced to collect the biomechanical data. Finally, the axial compression stiffness before and after fatigue test were calculated. There was no significant difference in the axial compression stiffness between pre- and post-fatigue test in each group ( P>0.05). However, the axial compression stiffness was significant higher in group A than that in groups B and C both before and after fatigue test ( P<0.05). No significant difference was found between group B and group C ( P>0.05). Self-made nickel-titanium shape memory alloy stepped plate is better than calcaneal plate and cannulated compression screws in axial load stiffness after being used to fix calcaneal osteotomy.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call