Abstract
The biomechanical performance of distal interphalangeal joint (DIPJ) fixation techniques in response to cyclic and load-to-failure bending loads is generally unknown. The purpose of this study is to assess the performance of 4 commonly used techniques for DIP arthrodesis. Phalanges were fixed by one of the following techniques: (1) 2 parallel Kirschner wires (K-wires), (2) 1 longitudinal K-wire and 1 oblique K-wire, (3) an Acutrak® 2 Micro headless compression screw, or (4) 90-90 wiring. The phalanges were loaded in cantilever bending on the palmar side of the phalange for 3000 cycles, then loaded-to-failure. The 90-90 wiring was significantly worse than the other 3 methods in 9 of 10 parameters analyzed. Both K-wire methods and the screw produced similar results in nearly all parameters. Given the lack of difference in biomechanical performance between K-wires and compression screws, consideration should be given to other factors such as cost and complication profiles when choosing an implant for DIPJ fusion.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.