Abstract

The present experimental study investigated biomechanical differences in methods of femoral lengthening using a monolateral external fixator only and using an external fixator over an intramedullary nail. Three materials, namely polyvinyl chloride rods, composite synthetic bone, and cadaver-bone, were tested using the MTS machine. We evaluated the differences of axial stiffness according to the presence of a nail or the numbers of half-pins (two or three half-pins) that were fixed at each side of osteotomy. The addition of a nail may increase the axial stiffness of the frame of monolateral external fixator for limb lengthening. Additionally, it is enough to distract the femur with fixing two half-pins at each side, when using the technique of lengthening over a nail.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.