Abstract

A biomagnetic measurement system was developed, suitable for the detection of magnetospinogram (MSG) and magnetocardiogram (MCG) signals from the dorsal surface of supine subjects. It is effective for noninvasively observing the electric activity of the spinal cord and/or heart. These biomagnetic signals are extremely weak, and magnetic flux sensors based on superconducting quantum interference devices (SQUIDs) are necessary to detect them. However, highly sensitive magnetic field measurement often suffers from ultra low-band circumstance noise mainly caused by transportation in urban areas. We applied reference sensors for monitoring the circumstance noise, and their outputs multiplied by appropriate weight coefficients were directly input to the feedback coil of a SQUID gradiometer. Synthesized in-phase components reduced the ultra low-band noise by approximately 90%. Both the MSG and MCG signals were successfully detected in a moderately magnetically shielded room. Even though the MCG signal band overlapped the ultra low-band noise, the signal-to-noise ratio was improved.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.