Abstract

This simple, rapid, sensitive kinetic bioluminescent method for the assay of bile acids in serum involves use of 3 alpha-hydroxysteroid dehydrogenase (EC 1.1.1.50) in combination with a new, commercially available NADH Monitoring Reagent (LKB-Wallac) containing a low activity of NADH:FMN-oxidoreductase and a high activity of bacterial luciferase. Interfering dehydrogenases in serum are inactivated in the test tube with trichloroacetic acid before the assay. The standard curve is linear for concentrations of bile acids up to about 300 mumol/L. With a sample volume of 20 microL, the detection limit is about 0.2 mumol/L. The within-run precision (CV) is about 10%, both at high and low concentrations of bile acids in serum. Correlation is good (r = 0.996) between results by this method and an enzymatic method based on spectrophotometry. However, the latter method is considerably less sensitive, and it is less precise at low concentrations of bile acids.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.