Abstract

Bioluminescence measurement significantly improved the accuracy, sensitivity, precision, and reliability of the current visual endpoint determination for the USP sterility test and eliminated the day 7 transfer/dilution step required for testing suspension products. Thirteen strains of bacteria and fungi (representing potential contaminants in sterile products), three pharmaceutical suspension products, and four media were used in the experiment. No interference from suspension products was encountered in the detection of microbial growth by the bioluminescence measurement. The poor fungal growth encountered was attributed to insufficient diffusion of oxygen into the medium and was circumvented by use of a large tube size (38 by 200 mm) or by vortexing the medium once during the 2-week incubation period. Bioluminescence measurement would facilitate automated handling of the sterility test endpoint readout operation. The optimum parameters of bioluminescence measurement for application in sterility testing were determined.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.