The aim of our study was to identify biological factors responsible for premature loosening of polyacetal hip stems. The results of histological analyses of the tissue around 11 total hip prostheses with loosened polyacetal femoral stems were compared to those obtained in a group of 11 total hip prostheses with loosened metal (CoCr) femoral components. A higher number of polymer wear particles surrounded by giant cells, more bone chips, and a more extensive necrosis were found around loosened polyacetal stems. Histomorphological characteristics of polyacetal wear particles containing BaSO(4) granules in the tissue around loosened polyacetal stems were described. Radiological evaluation of the wear of polyethylene cups suggested that elastic modulus of the stem had no influence on the wear of polyethylene cups. This study indicates that polyacetal wear particles have a great biological potential accelerating the process of loosening.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.