Abstract

A microbiological hood was designed to reduce the danger of airborne infection of laboratory workers. The hood uses absolute filters to deliver sterile air in a laminar flow to the work area. An air curtain across the hood opening permits easy access but separates the worker from aerosols produced in the hood, and protects material inside the hood from contamination by room air. Tests with bacterial and viral aerosols showed that the air curtain is at least 99.96% effective in preventing airborne particles from entering the work area.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.