Abstract
To investigate microbiological biofilm contamination of retrieved bone-anchored hearing aids. Nine failed, retrieved bone-anchored hearing aids and 16 internal screws were examined by scanning electron microscopy. A fixture from a failing implant, which had been removed and disassembled under aseptic conditions, was cultured. Finally, an internal screw from a new, unimplanted fixture was examined by scanning electron microscopy. Debris was seen on the fixture and abutment of all bone-anchored hearing aids, and on the heads of the 16 internal screws. On eight screws, biofilm extended down the shaft to the threads, where it was several micrometres thick. Culture of a failing fixture yielded staphylococcus. The new, unimplanted fixture internal screw showed evidence of scratching and metallic debris on the threads, which may interfere with close fitting of the screw and subsequently facilitate microleakage. There may be a link between internal microbial contamination and failure of bone-anchored hearing aids.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.