Abstract

The Howland current pump is a popular bioelectrical circuit, useful for delivering precise electrical currents. In applications requiring high precision delivery of alternating current to biological loads, the output impedance of the Howland is a critical figure of merit that limits the precision of the delivered current when the load changes. We explain the minimum operational amplifier requirements to meet a target precision over a wide bandwidth. We also discuss effective compensation strategies for achieving stability without sacrificing high frequency output impedance. A current source suitable for Electrical Impedance Tomography (EIT) was simulated using a SPICE model, and built to verify stable operation. This current source design had stable output impedance of 3.3 MΩ up to 200 kHz, which provides 80 dB precision for our EIT application. We conclude by noting the difficulty in measuring the output impedance, and advise verifying the plausibility of measurements against theoretical limitations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.