Abstract

The transverse intrafascicular multichannel electrode (TIME) is intended to be transversally implanted in the peripheral nerve and to selectively interface subsets of axons in different fascicles within the same nerve. Two versions of TIME (TIME-2, TIME-3) were designed and tested for biocompatibility and safety in the sciatic nerve of the rat. TIME-2 was implanted in two groups: one group had only an acute implant and the second group had chronic implantation for 2 months; a third group was also chronically implanted with the TIME-3 version, designed to avoid the mechanical traction produced by muscles motion. We evaluated the functional and morphological effects of either TIME-2 or TIME-3 implanted in the rat sciatic nerve for 2 months. The results of the study indicate that implantation of the TIME-2 and TIME-3 devices in the rat sciatic nerve did not cause significant axonal loss or demyelination, as evidenced by the functional and histological results. The results of this study indicate that the TIME-2 and TIME-3 designs are biocompatible and safe after chronic implantation in a small peripheral nerve, such as the rat sciatic nerve.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.