Abstract

A newly developed fiber optic micropressure sensor was evaluated for biocompatibility using the International Organization for Standardization (ISO) test standard 10993-6. The test material and an inert control (fused silica glass) were tested in New Zealand white rabbits. Four test specimens were implanted in the paravertebral muscles on one side of the spine about 2–5 cm from the mid-line and parallel to the spinal column. Similarly, four control specimens were implanted on the opposite side. The implantation periods were 1, 4, and 12 weeks to ensure a steady state biological tissue response. Four animals were tested at each time period. Macroscopic and microscopic observations were performed to compare the biological reactions between the test and control materials. There was an inflammatory reaction at 1 week which subsided at 4 weeks. There was fibrous tissue growth near the implant that also decreased over time. Most importantly, there was no significant difference in the biological response between the test and control materials. Therefore, we conclude that the pressure microsensor is biocompatible.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.