Abstract
Background/Aims: The use of adsorption cartridges for hemoperfusion (HP) is rapidly evolving. For these devices, the potential induced cytotoxicity is an important issue. The aim of this study was to investigate potential in vitro cytotoxic effects of different sorbent cartridges, HA130, HA230, HA330, HA380 (Jafron, China), on U937 monocytes. Methods: Monocytes were exposed to the sorbent material in static and dynamic manners. In static test, cell medium samples were collected after 24 h of incubation in the cartridges. In dynamic test, HP modality has been carried out and samples at 30, 60, 90, and 120 min were collected. Results: Compared to control samples, there was no evidence of increased necrosis or apoptosis in monocytes exposed to the cartridges both in the static and dynamic tests. Conclusion: Our in vitro testing suggests that HA cartridges carry an optimal level of biocompatibility and their use in HP is not associated with adverse reactions or signs of cytotoxicity.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.