Abstract
Near infrared spectroscopy (NIRS) can monitor both the redox status of Cytochrome c oxidase located in the mitochondria within the cell and the oxygenation of the blood in the tissue being monitored. Since the enzyme catalyzes more than 90% of oxygen utilization, it is the sink for the oxygen while the hemoglobin in the capillaries is the oxygen source. In order to evaluate the oxidative metabolic status of a tissue the optical data obtained from both molecules are commonly interpreted on the basis of test tube experiments with purified preparations. We are concerned that the validity of this practice may not have been tested sufficiently and raise four basic questions that have not yet been answered. Citing some examples of in vitro versus in vivo differences we conclude that more effort should be expended on the in vivo testing of the range of the signals, their natural variability, and the physiological and pathological meaning of their deviations from norm. © 1999 Society of Photo-Optical Instrumentation Engineers.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.