Abstract
Dopamine dissolved within egg homogenate was used as a model system to study the effects of electrode contamination and its subsequent reactivation through ultrasonically mediated in situ cleaning effects. The merits in conducting electroanalytical investigations under the influence of the ultrasonic field were also appraised. Maintaining the ultrasound field during oxidative measurements was found to yield hydrodynamic profiles that were linear over the concentration range 2-20 microM dopamine. The resulting sonolinear sweep voltammograms were compared with conventional rotating disk measurements, with the former found to provide significantly increased limiting currents that were attenuable through the manipulation of the field intensity. The problem of retaining selectivity in the presence of high concentrations of ascorbate was also assessed with the addition of cupric ion prior to commencing the measurements found to efficiently negate an otherwise substantive interference.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.