Abstract
A sensitive and selective LC-MS/MS method has been developed and validated for CDRI antidiabetic candidate S007-1261 in rat plasma using 16-dehydropregnenolone as an internal standard. The API 4000 triple quadrupole LC-MS/MS system was operated under multiple reaction monitoring mode using electrospray ionization technique in positive mode. The sample processing method involves 2-step liquid-liquid extraction using n-hexane as an extracting solvent. The analyte was chromatographed on RP 18, waters column (3.5 µm, 2.1 mm i.d. × 30 mm) with guard using acetonitrile and ammonium acetate buffer (pH 5.0, 10 mM) in 90:10 (v/v) composition at a flow rate of 0.40 mL min(-1). The chromatographic run time was 5.30 min. Calibration curve shows linearity over concentration range 1.56-200 ng mL(-1). The lower limit of detection was 0.39 ng mL(-1) and lower limit of quantitation was 1.56 ng mL(-1). The inter- and intra-day accuracy and precision were found to be within the assay variability limits as per US FDA guidelines. The absolute recovery of S007-1261 was found to be >90%. S007-1261 does not show any stability problems as it was stable at room temperature for 8 h. S007-1261 was also stable up to 3 freeze-thaw cycles and can be stored up to 30 days at -60 °C. The assay was successfully applied to both oral (40 mg kg(-1)) and intravenous (10 mg kg(-1)) pharmacokinetic studies in male Sprague-Dawley rats. The oral bioavailability of S007-1261 was found to be 33.61%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.