Abstract

Purpose: A rapid test for influenza viruses (Binax NOW®) was evaluated. Materials and Methods: In season-1, 35 respiratory samples were tested retrospectively; in season-2, 45 samples were tested prospectively [gold-standard: viral culture in season-1, culture+ reverse transcriptase-polymerase chain reaction (RT-PCR) in season-2]. Results: Sensitivity for Binax for influenza A was 59.3 and 0% in season-1 and -2, respectively. Sensitivity was low for influenza B (33.3% in season-1, 26.1% in season-2). Samples having low viral load were more likely to have a negative Binax test. Specificity of Binax was high (100 and 94.7% with influenza A and B, respectively). Conclusion: Sensitivity information provided in the kit insert does not always reflect post licensure performance in clinical settings.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.