Abstract
Dual-energy radiographic acquisition performed with a single pulse of x rays has been limited to use of stacked detectors such as photo-stimulable phosphor plates. In this study, a binary screen system is introduced that enables single-pulse dual-energy acquisition with nonstacked detectors such as charge-coupled devices (CCDs). Two x-ray phosphors with different K edges, designed to emit light of different wavelengths, were bound together in a single screen. Two CCD cameras, each sensitive to the respective wavelengths emitted by each phosphor, are proposed to be coupled to the binary screen. Optical isolation and detection of the optical emissions from the screen would enable simultaneous acquisition of both low- and high-energy images, which can then be used in dual-energy subtraction. Computer simulation results and experimentally acquired images suggest that the binary screen approach may be a viable alternative to stacked detector technology for dual-energy radiographic imaging.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.