Abstract

BackgroundTranscutaneous bilirubinometry is widely used to predict hyperbilirubinemia by using several devices. The aim of this study was to compare the predictive ability of BiliCheck vs JM-103 in identifying neonates not at risk of significant hyperbilirubinemia, putting the data obtained with the two instruments on our transcutaneous bilirubin nomogram built with the BiliCheck.MethodsTranscutaneous bilirubin (TcB) measurement was performed when jaundice appeared in newborn babies and/or just before discharge from the hospital. It was performed at the forehead with the two instruments within 5 minutes by two experienced neonatologists, each one blind to the value obtained by the other. Blood samples were drawn to obtain total serum bilirubin (TSB) levels soon after TcB measurements.ResultsA total of 627 paired-sample measurements were obtained from 298 newborn babies. Out of the total population studied, 16 newborn babies (5.4%) showed significant hyperbilirubinemia defined as TSB value >17 mg/dL, or as need for phototherapy treatment according to the AAP guidelines. TcB measurements showed false negative results in the first 60 hours of life using both devices. After the 60th hour of life, TcB measurements using both devices successfully predicted newborn babies not at risk of significant hyperbilirubinemia, being the JM-103 more reliable than BC because of fewer false positive results.ConclusionsOur study shows that both BC and JM-103 can exclude subsequent significant hyperbilirubinemia when the measurements are performed after the 60th hour of life. Nevertheless, the transcutaneous pre-discharge screening should be considered only as the first step, and it has to be followed by a follow-up through the first days after discharge.

Highlights

  • Transcutaneous bilirubinometry is widely used to predict hyperbilirubinemia by using several devices

  • The purpose of this study was to compare the predictive ability of BiliCheck vs JM-103 in identifying neonates not at risk of significant hyperbilirubinemia, putting the data obtained with the two instruments on our transcutaneous bilirubin nomogram built with the BiliCheck

  • Significant hyperbilirubinemia was defined as total serum bilirubin (TSB) value >17 mg/dL, or as need for phototherapy treatment according to the AAP guidelines

Read more

Summary

Introduction

Transcutaneous bilirubinometry is widely used to predict hyperbilirubinemia by using several devices. The aim of this study was to compare the predictive ability of BiliCheck vs JM-103 in identifying neonates not at risk of significant hyperbilirubinemia, putting the data obtained with the two instruments on our transcutaneous bilirubin nomogram built with the BiliCheck. Visual assessment of jaundice and risk factors evaluation did not give positive results in predicting subsequent hyperbilirubinemia [3,4], while plotting total serum bilirubin (TSB) measurements on hour-specific nomograms allowed the reduction in hospital readmission rate for severe hyperbilirubinemia (TSB > 30 mg/dl) [5,6,7,8].

Objectives
Methods
Results
Discussion
Conclusion

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.