Abstract

A novel setup for 23Na MRI, which allows bilateral imaging of the breast, is presented. For this purpose a figure-eight receive-only 23Na surface coil was developed. For our experiments on three samples with NaCl solutions of different sodium concentrations and two female subjects we used an asymmetric birdcage coil in transmit mode and the developed surface coil for receiving the signal at 3T. Imaging of the samples showed the applicability of the employed normalization method for measuring the distribution of sodium concentration. In a sample of concentration [Na+]=51mM we achieved SNR=70 at a nominal isotropic resolution of 2,5mm (TR=66ms, TE=0,6ms, TA=20min). Furthermore we showed that by means of this setup it is possible to quantify the sodium concentration in breast tissue (TSC) of a female subject with an accuracy of 23% (TR=150ms, TE=0,5ms, TA=45min).