Abstract
A novel setup for 23Na MRI, which allows bilateral imaging of the breast, is presented. For this purpose a figure-eight receive-only 23Na surface coil was developed. For our experiments on three samples with NaCl solutions of different sodium concentrations and two female subjects we used an asymmetric birdcage coil in transmit mode and the developed surface coil for receiving the signal at 3T. Imaging of the samples showed the applicability of the employed normalization method for measuring the distribution of sodium concentration. In a sample of concentration [Na+]=51mM we achieved SNR=70 at a nominal isotropic resolution of 2,5mm (TR=66ms, TE=0,6ms, TA=20min). Furthermore we showed that by means of this setup it is possible to quantify the sodium concentration in breast tissue (TSC) of a female subject with an accuracy of 23% (TR=150ms, TE=0,5ms, TA=45min).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.