Abstract
A comparison of the 10 V Josephson array voltage standard of the Physikalisch-Technische Bundesanstalt (PTB), Germany, was made with that of the Primarni Elektromagnetski Laboratorij (PEL), Croatia, in April 2007. The results are in very good agreement and the measured overall relative combined standard uncertainty is better than 1 part in 1010.Main text.To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.The final report has been peer-reviewed and approved for publication by EURAMET, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.