Abstract

Bilateral comparisons of the calibration of low nominal capacity torque measuring devices (TMDs) were conducted between NMIJ and PTB for the first time. A 10Nm deadweight torque standard machine (TSM) newly developed at NMIJ was compared with PTB’s two TSMs with nominal capacities of 1Nm and 1kNm, in the range from 0.1Nm to 10Nm. The transfer measurement devices used had nominal capacities of 1Nm and 10Nm. The comparisons were made using two calibration procedures: one based on the CCM.T-K2 Key Comparison procedure and the other according to the daily calibration procedure adopted at each institute. As a result, the torque realized by the TSM at NMIJ was shown to be equivalent to that realized by the two TSMs at PTB. In addition, it was confirmed that the daily calibration procedures of both institutes yielded consistent results in the calibration of low nominal capacity TMDs.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.