Abstract

A bilateral comparison has been made between the air-kerma standards of the IAEA and the VNIIM in the medium-energy x-ray range. The results show the standards to be in agreement at the level of the stated standard uncertainty for the comparison of 4.1 parts in 103. The comparison result, based on the calibration coefficients for a transfer chamber, is expressed as a ratio of the IAEA and the VNIIM standards for air kerma, suitable for entry in the BIPM key comparison database.Main text.To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.The final report has been peer-reviewed and approved for publication by the CCRI, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).